Imaging performance
|
|
Resolution
|
640x480 pixels
|
|
Focal Plane array (FPA)
|
Uncooled microbolometer 640x480 pixels, latest generaton with 17 µm pitch
|
|
Spectral range
|
7.8 to 14 µm
|
|
Field of view / min focus distance
|
25° x 19° / 0.25 m
|
|
Spatial resolution (IFOV)
|
0.68 mrad
|
|
Image frequency
|
30 Hz
|
|
Focus
|
Manual / autofocus
|
|
|
|
Image presentation
|
|
Display
|
4.3" superbright touchscreen LCD 800x480 pixels
|
|
Image modes
|
IR-image with selected color scale, Full color visual, Picture in Picture (Resizable and movable IR-area), Thermal Fusion (Threshold above, below and interval), thumbnail gallery
|
|
Manual image adjustments
|
Level/span/max/min
|
|
Automatic image adjustments, continuous or manual activation
|
Standard or based on histogram from image content
|
|
Automatic image adjustment with locked scale
|
Lock max, min or span
|
|
|
|
Measurement
|
|
Accuracy
|
± 2 °C or ± 2% of reading
|
|
|
|
Measurement analysis
|
|
Spotmeter
|
10
|
|
Area
|
5 Max/Min/Average value within box or circle
|
|
Automatic hot/cold detection
|
Max/Min temp. value and position shown within box, circle or on a line
|
|
Isotherm
|
Detect high/low temperature/interval
|
|
Difference temperature
|
Difference between any two measurement functions or any measurement function and a reference temperature
|
|
Reference temperature function
|
Manually set
|
|
Emissivity correction
|
Variable from 0.01 to 1.0 or selected from materials list
|
|
Measurement corrections
|
Reflected temperature, optics transmission and atmospheric transmission
|
|
External windows correction
|
Automatic based on inputs of window temperature and transmission
|
|
|
|
Set-up
|
|
Image controls
|
Palettes (Arctic, Gray, Iron, Lava, Rainbow and Rainbow HC), image adjustment (auto/manual)
|
|
Set-up controls
|
Local adaptation of units, language, date and time formats; automatic shutdown , display intensity
|
|
Configure information to be shown in image
|
Yes
|
|
Programmable button
|
Yes
|
|
|
|
Image storage
|
|
Type
|
IR/visual images; simultaneous storage of visual and IR images
|
|
Format
|
Standard JPEG - including measurement data on SD memory card
|
|
|
|
Digital camera
|
|
Built-in digital camera
|
5 Mpixel incl. lamps
|
|
|
|
Laser LocatIR
|
|
Laser
|
Semiconductor AlGaInP diode laser, Class 2 - position is displayed on the IR image
|
|
Laser alignment
|
Laser position shown on IR-image
|
|
|
|
Image annotation
|
|
Voice
|
60 seconds via Bluetooth®
|
|
Text
|
Text from predefined list or soft keyboard on touch screen
|
|
Sketch
|
A sketch drawn on touch screen is automatically saved with image
|
|
Meterlink
|
Wireless connection to: Extech Moisture meter MO297 or Extech clamp meter EX845
|
|
|
|
Report generation
|
|
Instant Report in camera
|
Automatic generation of PDF report based on selected images direct in camera
|
|
|
|
Video streaming /recording
|
|
Non radiometric IR-video streaming
|
MPEG 4 streaming to PC using USB
|
|
Video recording in camera
|
Non-radiometric IR video/visual video, MPEG4 to SD-card.
|
|
WiFi
|
Wireless streaming of non-radiometric IR-video, MPEG4
|
|
|
|
Update of camera
|
|
Automatic update of camera to latest version
|
Automatic update of camera from PC running FLIR Tools
|
|
|
|
Data communication interfaces
|
|
Interfaces
|
USB-mini, USB-A, Bluetooth®, WiFi, DVI video
|
|
USB
|
USB-A: Connect external USB device - USB-mini-B: Data transfer to and from PC / Streaming MPEG 4
|
|
WiFi
|
Connects directly to Ipad/Iphone for image transfer
|
|
|
|
Power
|
|
Battery type
|
Lithium-Ion (field replaceable)
|
|
Battery operating time
|
> 2.5 hours at 25°
|
|
Charging system
|
In camera, AC adaptor, 2-bay charger or 12 V from a vehicle
|
|
Power management
|
Automatic shutdown and sleep mode (user selectable)
|
|
AC operation
|
AC adaptor, 90-260 V AC, 50/60 Hz
|
|
Adaptor voltage
|
12 Volt VDC out
|
|
|
|
Environmental specifications
|
|
Operating temperature range
|
-15 to +50°C
|
|
Storage temperature range
|
-40 °C to +70 °C
|
|
Humidity, operating and storage, non-condensing
|
IEC 60068-2-30 /24 h, 95% relative humidity +25 °C to +40 °C
|
|
Encapsulation
|
IP 54, IEC 60529
|
|
Bump, Operational
|
25G, IEC 60068-2-29
|
|
Vibration, Operational
|
2G, IEC 60068-2-6
|
|
EMC, emission
|
EN 61000-6-3
|
|
EMC, immunity
|
EN 61000-6-2
|
|
|
|
Physical characteristics
|
|
Camera weight incl. battery
|
1.3 kg
|
|
Camera size (L x W x H)
|
143 x 195 x 95 mm
|
|
Tripod Mounting
|
1/4" - 20
|
|
|
|
Standard package
|
|
FLIR T620 / T640: Hard transport case, thermal imaging camera with lens, Battery (2), Battery charger, Large eyecap, Tripod adaptor, Neck strap, Lens cap, Bluetooth® headset, Calibration certificate, FLIR Tools™ PC software CD-ROM, Memory card with adaptor, Power supply incl. multiplugs, Printed Getting Started Guide, Printed Important Information Guide, USB cable, User documentation CD-ROM, HDMI cable (2), Warranty extension card or Registration card
|